Recall Z-1865-2011 ยท IXI Block, Beam-Shaping, Radiation Therapy

Xoft Inc.

Axxent Flexishield Mini To shape the beam from a low energy radiation therapy source up to 50kVp.

DateApr 1, 2011 (initiated Feb 3, 2011)
StatusTerminated
ReasonXoft Axxent Flexishield Mini product may may shed particles identified as tungsten, and may look like suspicious calcifications on follow up scans if the product was used during Intraoperative radiation therapy.
Root causeDevice Design
ActionXoft, Inc. sent an Urgent Field Removal Action letter dated February 3, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to stop using and return all units and lots of Flexishield Mini Catalog Number: F5300 in their inventory. Customers were instructed to fill out the attached form and fax to 866-222-3404, then call Customers Service toll free directly at 877-963-8327 to make arrangements to return the product and receive a credit if applicable. Customers were asked to complete and retu
Quantity165 units
DistributionNationwide Distribution including AL, CA, CO, FL, IL, MI, MO, MS, NJ, NY, OK, PA, TN, and WI.