Recall Z-2041-2011 ยท EHD Unit, X-Ray, Extraoral With Timer

Gendex Dental Systems

Orthoralix 9200 DDE Panoramic and Cephalometric Dental X-Ray System; Extraoral x-ray unit with timer; Gendex Dental Systems, Des Plaines, IL 60018; Part numbers 110-0199G2, 110-0199G4, 110-0234G1 and 110-0234G3. The intended use of the device is to be used as an extraoral source of x-rays for imaging of the dento-maxillo-facial area.

DateApr 22, 2011 (initiated Feb 4, 2011)
StatusTerminated
ReasonThere is an anomaly in the Orthoralix firmware which can cause a message corruption which will distort the patient image being transferred from the X-Ray system to the PC. This image distortion/corruption (distorted, fragmented, flipped, negative, reversed or had lines across them) is most likely to occur during periods of high network traffic and is significant enough that it is clear and obviou
Root causeOther
ActionGendex sent a recall letter to their accounts on February 4, 2011 via first class mail, stating that an anomaly in the Orthoralix firmware can cause a message corruption which will distort the patient image being transferred from the X-Ray system to the PC. This image distortion/corruption (distorted, fragmented, flipped, negative, reversed or had lines across them) is most likely to occur during periods of high network traffic and is significant enough that it is clear and obvious these images cannot be used for diagnostic purposes. The letter stated that resending the image from the X-Ray
QuantityTotal 6,835 units of all products
DistributionWorldwide Distribution--United States, including Puerto Rico, Australia, Canada, Chile, Colombia, Germany, Hong Kong, India, Italy, Korea, Mexico, New Zealand, Peru, Singapore, Taiwan and Vietnam.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032355ORTHORALIX 9200 DDE, ORTHORALIX 9200 DDE PLUS CEPH, MODELS 5105 0161WW, 5105 0181WW
Dentsply Gendex, Sep 25, 2003