Recall Z-2897-2011 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Toshiba American Medical Systems Inc

Aplio Artida Ultrasound Diagnostic System SSH-880CV. Toshiba American Medical Systems, Inc. A diagnostic ultrasound system.

DateJul 29, 2011 (initiated Oct 1, 2010)
StatusTerminated
ReasonToshiba America Medical Systems, Inc. is initiating a field correction on the Aplio Artida Ultrasound Diagnostic System because the panel operation for the Aplio Artida may become disabled (i.e., system may hang up) when processing to display patient information (heart rate, time, VCP counter, ect.).
Root causeSoftware design
ActionToshiba America Medical Systems, Inc. sent out an "Urgent Medical Device Correction" dated December 21, 2010 to all affected customers. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were notified that Toshiba will issue a field modification to correct the software, and if the problem occurs, they were instructed to shutdown the system holding down the power supply button for 5 seconds for 5 seconds and then restart the system. A form to send back to Toshiba is included. For questions on this recall call (800) 421-1968
Quantity46 units
DistributionWorldwide Distribution: (USA) Nationwide distribution including the states of MD, OH, PA, NY, MA, GA, AL, TX, IL, WI, IA, MN, KY, MT, OR, LA, and CA; and the country of Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090158APLIO ARTIDA V2.0 SSH-880CV
Toshiba America Medical Systems, In.C, Feb 6, 2009