ClearanceBrief
Recall Z-2030-2011 ยท GWM Device, Monitoring, Intracranial Pressure

Codman & Shurtleff, Inc.

Codman Metal Disposable ICP Kit Product Code: 80-1190 This is a single use, disposable drill kit designed for gaining access to the cranium for therapeutic and/or diagnostic procedures

DateApr 20, 2011 (initiated Nov 30, 2010)
StatusTerminated
ReasonPackaging defect may compromise sterility of device
Root causePackage design/selection
ActionThe firm, Codman, sent an "Important Notice Urgent- Voluntary Product Recall Notification" letter dated November 30, 2010 to its customers via FedX. On January 7, 2011, Codman expanded the recall to four additional product codes.On January 31, 2011, the firm issued a supplemental recall letter and identified one additional product code: 80-11197. The letters described the product, problem and actions to be taken. The customers were instructed to identify the unused affected products in their inventory and follow the recall instructions provided: 1) If you have NO product to return, please advi
Quantity1671 units
DistributionWorldwide distribution: USA including states of :AL, FL,GA, KY, MA, MI, MN, MS, OR, PA, RI, UT,WI and WV; and countries including: Argentina, Australia, Canada, Bahamas , Belgium, Dubai, Japan, Madrid, Mexico, New Zealand, Saudia Arabia, UAE, Uruguay, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.