Recall Z-0833-2012 ยท RCE Cabinet X-Ray, Industrial
GE Healthcare, LLC
GE eXplore Locus Micro CT cabinet x-ray system model RS9-80. CT Scanner for laboratory mice or other research.
| Date | Feb 17, 2012 (initiated Jan 31, 2011) |
| Status | Terminated |
| Reason | Product fails to comply with the federal performance standard for cabinet x-ray systems (Title 21 CFR 1020.40(c)(1)(i)). Specifically, the x-ray emission limit of 0.5 milliroentgen in one hour could be exceed. Highest emission measured was 3 mR/hr. |
| Root cause | Radiation Control for Health and Safety Act |
| Action | GE Healthcare will notify purchasers of the problem, the affected product, instructions for users, and specify the maximum emission rate measured. GE Healthcare will bring defective systems into compliance free of charge. This will be done via Field Modification Instructions IFMI) 11101 which is scheduled for deployment on March 1, 2011 with an estimated completion date of July 1, 2011. |
| Quantity | 21 |
| Distribution | Worldwide Distribution |
| FDA record | Open on fda.gov |