Recall Z-0833-2012 ยท RCE Cabinet X-Ray, Industrial

GE Healthcare, LLC

GE eXplore Locus Micro CT cabinet x-ray system model RS9-80. CT Scanner for laboratory mice or other research.

DateFeb 17, 2012 (initiated Jan 31, 2011)
StatusTerminated
ReasonProduct fails to comply with the federal performance standard for cabinet x-ray systems (Title 21 CFR 1020.40(c)(1)(i)). Specifically, the x-ray emission limit of 0.5 milliroentgen in one hour could be exceed. Highest emission measured was 3 mR/hr.
Root causeRadiation Control for Health and Safety Act
ActionGE Healthcare will notify purchasers of the problem, the affected product, instructions for users, and specify the maximum emission rate measured. GE Healthcare will bring defective systems into compliance free of charge. This will be done via Field Modification Instructions IFMI) 11101 which is scheduled for deployment on March 1, 2011 with an estimated completion date of July 1, 2011.
Quantity21
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.