Recall Z-1493-2011 ยท DRX Electrode, Electrocardiograph

ConMed Corporation

Neotrode Wire Attached, Catalog Number: 1731-003; Neotrode II Wire Attached, Catalog Number: 1741-003; Softrace Neonatal Wire Attached, Catalog Number: 2321-003; Softrace Pediatric Wire Attached, Catalog Number: 2331-003; Softrace Limb Band, Catalog Number: 2310-003; TRU-LINK ECG Electrode (Neonatal Pre-Wired), Catalog Number: 685-0037-00; for use with electrocardiographic monitoring equipment to

DateMar 2, 2011 (initiated Sep 23, 2010)
StatusTerminated
ReasonPoor signal or loss of signal. Certain recalled devices have a loose wire attachment that may lead to intermittent or poor signal quality.
Root causeNonconforming Material/Component
ActionConMed Corporation distributed an Urgent Device Recall notification letter on Tru-Link ECG Electrode (Neonatal Pre-wired) dated September 23, 2010. ConMed Corp. has advised to cease the use of the devices immediately and conducted a recall to the user level. Direct consignees were asked to contact sub-accounts if recalled devices were further distributed and notified them of such recall. Recalled devices were to return to ConMed Corp. at 525 French Road, Utica, NY 13502 with completed Attachment II to indicate the amount returned. Customers with no recalled devices were asked to fill out A
Quantity9,009,000 Pediatric/Neonatal ECG Electrodes with Attached Wire
DistributionWorldwide distribution, including USA, Dubai, Ecuador, Egypt, Germany, Holland, Indonesia, Israel, Italy, Jamaica, Kenya, Korea, Lithuania, Malaysia, Mexico, New Zealand, Norway, Qatar, Saudi Arabia, Singapore, South Africa, Sweden, Switzerland, US Virgin Islands, Canada and Belgium.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K901555MODEL 1730 NEOTRODE WITH ATTACHED LEADWIRE
Medtronic Vascular, Jul 18, 1990