Recall Z-1435-2011 ยท LLZ System, Image Processing, Radiological

AGFA Corp.

IMPAX CV Results Manager/Results Manager Administration Tool The IMPAX Cardiovascular suite is a cardiovascular information system providing image archiving and image display and modality study-specific structured reporting. The suite is comprised of modular software that runs on industry standard personal computers and servers, utilizing Windows 2000/2003/XP operating systems. Image data storage

DateFeb 25, 2011 (initiated Apr 15, 2009)
StatusTerminated
ReasonUser could modify a measurement's units and that change would not be reflected in echocardiography formulas used within the Adult echocardiography RM.
Root causeOther
ActionThe firm, Agfa Healthcare, sent individual letters dating from April 15, 2009 to August 26, 2009, to the customers. The letters described the product, problem and actions to be taken. Agfa provided the affected customer with a complete list of their impacted patients and the measurements. Agfa also scheduled unit corrections with their customers and received acknowledgement of the letters at the time of scheduling the unit corrections. Service Bulletin DIS026.09E was issued January 2009 to inform users that Agfa will no longer allow customers to request unit changes for measurements that hav
Quantity6 devices
DistributionNationwide distribution: USA including states of: IL, MT, NY, OR and SC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050228AGFA HEARTLAB CARDIOVASCULAR
Heartlab, Inc., Feb 18, 2005