Recall Z-1431-2011 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Healthcare, LLC

8) Signa 3.0T Infinity with EXCITE (K030874 GE 3.0T Signa Infinity TwinSpeed with EXCITE MR System) Signa 3.0T with EXCITE (K030874 GE 3.0T Signa Infinity TwinSpeed with EXCITE MR System) The GE Signa@ Infinity line of products are whole body magnetic resonance scanners designed to support high resolution, high signal-to-noise ratio and short scan times. The Signa@ Infinity System is indicated

DateFeb 24, 2011 (initiated Nov 9, 2010)
StatusTerminated
ReasonGE Healthcare has become aware of a potential issue associated with the gradient cables being inadvertently swapped while servicing the gradients on some GE Healthcare MR Systems that may impact patient safety. If the gradient cables are inadvertently swapped while servicing the gradients, and the Service Engineer does not perform the geometry check as required in the service procedure, images m
Root causeNonconforming Material/Component
ActionOn November 9, 2010, the firm, GE Healthcare, sent an "Urgent Medical Device Correction" letter dated November 5, 2010 to all Consignees/Customers. The letter described the product, problem and action to be taken by the firm. GE instructed the customers to contact their local GEHC support Engineer, if they have any questions about the image orientation, or the procedures for geometry checks on thier system. Additionally, a GE Healthcare representative will install a device, which will reduce the potiential for the gradient cables to be swapped. Contact your local GEHC Support Engineer if yo
Quantity73
DistributionWorldwide distribution: USA including DC and PR; and countries including: VIET NAM, VENEZUELA, URUGUAY, UNITED KINGDOM, UNITED ARAB EMIRATES, UKRAINE, TURKEY, TUNISIA, FORMER YUGOSLAV REPUBLIC OF MACEDONIA, THAILAND, TAIWAN, SYRIAN ARAB REPUBLIC, SWITZERLAND, SWEDEN, SPAIN, SOUTH AFRICA, SLOVENIA, S
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030874GE 3.0T SIGNA INFINITY TWINSPEED WITH EXCITE MR SYSTEM
General Electric Co., Mar 28, 2003