Recall Z-1655-2011 ยท IZI System, X-Ray, Angiographic

Philips Medical Systems North America Co. Phillips

Philips Allura Integris Systems with OMCP Generator, x-ray diagnostic system. FCO72200185.

DateMar 16, 2011 (initiated Oct 7, 2010)
StatusTerminated
ReasonA wire in the connector of the Generator Grid Switch Supervisor box might short-circuit between 12V power and ground, and it will cause a System breakdown.
Root causeNonconforming Material/Component
ActionOn 10/15/2010, Philips began sending out the URGENT-Device Correction Notice Failure Causing the Generator to Stop Working notification letter to their consignees. The consignees were informed that a short circuit may occur due to a faulty wire connection in the Supervisor Box Connector in the Generator, and it will cause the generator stops working. When the system fails during a critical interventional case, this may cause risk to patients (no Fluoro available). The systems involved in this recall are Allura Xper with the Velara Generator, Allura Integris with OMCP Generator, MultiDiagnost
Quantity1078 units distributed in US (1102 +24-48=1078)
DistributionDevices were distributed through out the US.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K984545RELEASE 2 OF PHILIPS INTEGRIS SERIES (V5000, BV5000, H5000, BH5000, HM2000) SYSTEMS
Philips Medical Systems North America, Inc., Feb 25, 1999