ClearanceBrief
Recall Z-1421-2011 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Siemens Medical Solutions USA, Inc

AXIOM SIRESKOP SD, SIRESKOP SX, AXIOM LUMINOS System, X-Ray, Fluoroscopic, image-intensified; solid state x-ray imager (flat panel/digital imager)

DateFeb 23, 2011 (initiated Jan 14, 2011)
StatusTerminated
ReasonFirm became aware of a potential malfunction and possible hazard to patients in the event the tilt buttons on the OPTI Grip become stuck under the housing
Root causeOther
ActionSiemens Medical Solutions USA, Inc issued a Customer Safety Advisory Notice via Update Instruction XP005/10/S to affected customers. This letter informs customers of the potential malfunction and of possible danger for patients, users or other persons. A future Update Instruction will be released for the replacement of the tilt buttons on all OPTI Grip handles. Customers were instructed to forward the safety notice to any new owners of this device and to provide Siemens Medical with names of the new customers. For any questions regarding this recall call 610-448-3237.
Quantity1323
DistributionNationwide Distribution including AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051602AXIOM SIRESKOP SD
Siemens Medical Systems, Inc., Jul 7, 2005
K062623AXIOM LUMINOS DRF
Siemens Medical Solutions USA, Inc., Aug 22, 2007
K951358SIRESKOP SX
Siemens Medical Solutions USA, Inc., May 15, 1995