Recall Z-1731-2011 ยท GKZ Counter, Differential Cell

Beckman Coulter Inc.

Coulter MAXM Hematology Analyzer and MAXM Hematology Analyzer with Autoloader. Hematology Analyzers; Part Number: 6604569, 6604687, 6705995, 6705996, 6705997, 6705998, 6705999, and 6706000. Quantitative, automated, differential cell counters for in vitro diagnostic use.

DateMar 18, 2011 (initiated Dec 14, 2010)
StatusTerminated
ReasonThe recall was initiated because Beckman Coulter has confirmed the Coulter MAXM, MAXM AL, HmX and HmX AL Analyzers omit the tilde (~) character when the tilde is used as part of the Sample 10 within a barcode label scanned by the primary mode barcode reader. There is a potential for Specimen or Patient misidentification to occur.
Root causeOther
ActionThe recall communication was initiated on 12/14/2010 with Beckman Coulter forwarding an Urgent: Product Corrective Action (PCA) letter with attached PCA Response Form (via US Postal Service for US Customers) to all customers who purchased the COULTER HmX Hematology Analyzer, COULTER HmX Hematology Analyzer with Autoloader, COULTER MAXM Analyzer, COULTER MAXM Hematology Analyzer with Autoloader. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed not to use the tilde (~) character in Specimen or Patient identifier
Quantity4288 units total
DistributionWorldwide Distribution -- US, Algeria, Andorra, Antigua and Barbuda, Australia, Austria, Bahrain, Bangladesh, Bermuda, Bosnia and Herzegovina, Botswana, Brunei Darussalam, Bulgaria, Burkina Faso, Burundi, Canada, Cayman Islands, China, Congo, Cote d'Ivoire, Croatia, Czech Republic, Djibouti, Egypt,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K911385COULTER(R) MH010
Coulter Electronics, Inc., Jun 17, 1991