Recall Z-1863-2011 ยท JJX Single (Specified) Analyte Controls (Assayed And Unassayed)

Abbott Point Of Care Inc.

Cardiac Marker Control Level 1, List Number 136601-06F12-02; Cardiac Marker Control Level 2, List number 136602/06F13-02; and Cardiac Marker Control Level 3, List number 136603/06F14-02. Used to verify the integrity of newly received i-STAT cTnI cartridges.

DateApr 1, 2011 (initiated Dec 17, 2010)
StatusTerminated
ReasonPresence of CO2 in the head space of the vial can result in the generation of high results in Cardiac Marker Control Level 1, 2, and 3 which is used along with the Cardiac Marker Control Calibration Verification Set.
Root causeDevice Design
ActionUrgent Recall Notice letters with business reply cards were sent Fed ex on December 17, 2010 to all accounts. International notifications were sent on December 21, 2010. The letter identified the affected product and lot numbers along with a background explanation of the issue. The firm recommends that customers remove the cap from the vial; squeeze the vial 10 times, prior to thawing; replace cap on the vial; and then follow the remaining steps for use as outlines in section 14-12 of the i-STAT 1 System Manual. Customers are to complete and return the Business Reply card to acknowledge r
Quantity2360 units total
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031873I-STAT CARDIAC MARKERS CONTROL LEVEL 1, 2, AND 3; CARDIAC MARKERS CALIBRATION VERIFICATION CONTROL SET
I-Stat Corporation, Oct 7, 2003