ClearanceBrief
Recall Z-1178-2011 ยท FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days

Cook, Inc.

Spectrum Central Venous Catheter Tray, Minocycline/Rifampin Antibiotic Impregnated, Triple Lumen Polyurethane, 9.0 Fr/ 20 cm, C-UTLMY-901J-RSC-ABRM-FST-RD, G49005, Sterile, Cook Medical, Cook Incorporated, Bloomington, IN.

DateFeb 23, 2011 (initiated Oct 4, 2010)
StatusTerminated
ReasonCook, Inc. is recalling Single, Double, Triple, and Five-Lumen Central Venous Catheter Trays AND Single and Double Lumen PICC Peripherally Inserted Central Venous Catheter Trays. These kits and trays contain 0.9% Sodium Chloride Injection USP 5 mL fill in 6 mL flush syringes that have been recalled by Excelsior Medical. During a routine test for syringe sterility, a possible integrity failure wa
Root causeNonconforming Material/Component
ActionCook, Inc. sent URGENT DEVICE RECALLS letters to its customers dated 10/4/2010. The letters asked customers to examine their inventory and notify Cook for product return. The letter included a response form.
Quantity51,473 total kits/trays
DistributionWorldwide distribution: USA, Chile, Colombia, Panama, Barbados, Canada, Guatemala, and Saudi Arabia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081113COOK AND COOK SPECTRUM CENTRAL VENOUS CATHETERS
Cook, Inc., Jul 30, 2008