Recall Z-1252-2011 ยท IYE Accelerator, Linear, Medical

Elekta, Inc.

Elekta Synergy XVI Intended use: Radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner.

DateFeb 15, 2011 (initiated Jan 5, 2011)
StatusTerminated
ReasonFlexmap calibration did not cover the full 360 degrees of gantry rotation (0 to 355 degrees) and that this caused an isocenter displacement with 2D PlanarView images required at 180 to -180.
Root causeOther
ActionThe firm, Elekta Limited, sent an "Important Notice A344" dated January 5, 2011 to all consignees/customers. The notice described the product, problem and action to be taken by the customers. The customers were instructed to follow any instructions or recommendations covered in the notice. The notice provides suggestions on how to check and prevent this situation from occurring; it instructs the User to file the document in the Important Notice section of the appropriate User Manual; to complete a FlexMap Check each time they acquire a new FlexMap and amend their local procedures to include
Quantity223 units
DistributionNationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051932ELEKTA SYNERGY; ELEKTA SYNERGY S AND XVIR3.5
Elekta, Inc., Aug 12, 2005