Recall Z-1644-2011 ยท FRN Pump, Infusion

Hospira Inc.

Plum A+ Single Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 11971 b) list 11973 - software version 10.3 c) list 12391 - software version 11.3 d) list 20679 - Hospira MedNet Software e) list 20792 - Driver The Plum A+ dual-line volumetric infusion system designed to meet the growing demand for hospital-wide, as well

DateMar 14, 2011 (initiated Feb 14, 2011)
StatusTerminated
ReasonHospira has received customer reports of the Plum A+ infusion pumps with no audible alarm conditions, which have been associated with failure of the piezoelectric assembly due to component quality issues.
Root causeNonconforming Material/Component
ActionThe firm, Hospira, sent "URGENT DEVICE RECALL" letters dated February 14, 2011 to their customers on the same date. The letter described the product, problem and action to be taken by the customers. The Customer/Clinicians were instructed to weigh the risk associated with continued use of the device versus removal from service. If they elect to continue the devices, they were provided with instructions to perform alarm tests prior to each clinical use (during their pump cleaning process). If the alarm is not audible, they were instructed to discontinue use of the pump and contact Hospira Globa
Quantity144,477 pumps
DistributionWorldwide distribution: United States including the U.S. Virgin Islands and Puerto Rico, and countries including: Argentina, Australia, Bahamas, Barbados, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011442ABBOTT PLUM A+INFUSION PUMP, MODEL 11971
Abbott Laboratories, Jun 5, 2001
K042081PLUM A+ INFUSION SYSTEM WITH HOSPIRA MEDNET SOFTWARE , AND PLUM A+3 INFUSION SYSTEM WITH HOSPIRA MEDNET SOFTWARE
Hospira, Inc., Aug 24, 2004
K070398HOSPIRA PLUM A+ INFUSION PUMP, MODEL 12391
Hospira, Inc., Apr 24, 2007
K982159ABBOTT PLUM A+ INFUSION PUMP
Abbott Laboratories, Jan 12, 1999