Recall Z-1602-2011 ยท FRN Pump, Infusion

Hospira Inc.

Hospira Plum XL3 Micro/Macro with DataPort Infusion Pump, Triple Channel, with a 0.1-999 mL/hr flow rate range; list number 11855

DateMar 8, 2011 (initiated Feb 14, 2011)
StatusTerminated
ReasonHospira has received reports of the Plum XL, XLM and XLD Infusion Pumps with no audible alarm at the low audio level setting.
Root causeLabeling design
ActionHospira sent Urgent Device Recall letters dated February 14, 2011 to their customers on the same date, informing them that Hospira had received complaints of the Plum XL, XLM and XLD infusion pumps in which the audible alarm failed. If the audible alarm fails and the user does not notice the visual alert, the user may not be aware of the change in pump status such as air-in-line or occlusion. This may result in a delay or interruption of therapy which may result in serious injury and/or death. Clinicians were instructed to weigh the risk associated with continued use of the device versus
Quantity1,291 pumps
DistributionWorldwide distribution, including United States, the U.S. Virgin Islands, Puerto Rico, Argentina, Australia, Bahamas, Bahrain, Barbados, Belgium, Bosnia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Ecuador, EI Salvador, Finland, France, Greece, Guatemala, Hong Kong, Hungar