Recall Z-1395-2011 ยท GTY Antigens, All Groups, Streptococcus Spp.

Biosite Inc Dba Innovacon Inc.

SP Brand Strep A Cassette; Part Number: B1077-26; Lot #: STA0010037 The SP Brand Strep A Rapid Test Device is a rapid chromatographic immunoassay for the qualitative detection of Strep A antigen from throat swab specimens to aid in the diagnosis of Group A Streptococcal infection. This test is indicated for professional and point of care use only.

DateFeb 22, 2011 (initiated Jan 12, 2011)
StatusTerminated
ReasonThe recall was initiated after the Investigation of customer-reported complaints, testing of market retention and returned product, It was determined that this lot of product has demonstrated an elevated Incidence of non-specific binding, potentially causing false positive results. A false positive test result could lead to the administration of antibiotics that are unnecessary. This exposes the
Root causePending
ActionThe firm, Alere San Diego, sent a "URGENT MEDICAL DEVICE RECALL" letter dated January 17, 2011. The letter described the product, problem and action to be taken by customers. The customers were instructed to: (1) Immediately discontinue use of the affected lot (if applicable) and inventory remaining full or partial kits at their site (if applicable); (2) Destroy any remaining inventory of this lot and document destruction per their site requirements (if applicable); (3) Complete and FAX the enclosed Verification Form within 10 days to confirm their receipt of this notice and to receive repla
Quantity814 kits/24,420 devices
DistributionNationwide distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031784ACON STREP A RAPID TEST DEVICE
ACON Laboratories, Inc., Jul 14, 2003