Recall Z-1604-2011 ยท IXL Device, Spot-Film

Philips Medical Systems North America Co. Phillips

Philips BV Endura, Rel. 2, 718074. Mobile C-Arm X-Ray systems offering Radiographic and Fluoroscopic techniques in a wide variety of applications.

DateMar 8, 2011 (initiated Dec 14, 2010)
StatusTerminated
ReasonThe fixing strap of the belt holding the CBX Laser to the Image Intensifier may be detached during a procedure when used with the BV Libra and the BV Pulsera/Endura Systems.
Root causeDevice Design
ActionOn 01/06/11, Philips sent an URGENT Device Correction Notice, dated 01/05/2011, to their consignees. The consignees were informed that the fixing strap of the Laser Aiming Device Image Intensifier may inadvertently detach when used with BV Libra, BV Endura Rel. 2, and BV Pulsera Rel 2.3 Mobile X-Ray systems. Specifically the draw latch of the fixing strap holding the CBX Laser to the Image Intensified can be accidentally released during a procedure when used with the above Mobile X-Ray systems. While waiting for Philips to implement their corrective action, the consignees are advised to
Quantity10 units
DistributionNationwide Distribution -- AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Washington, DC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010435PHILIPS BV PULSERA/ENDURA
Philips Medical Systems, Inc., Mar 12, 2001