ClearanceBrief
Recall Z-0960-2011 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Sorin Group USA, Inc.

S5 Control Panel for 2 Mast Roller Pumps 85, Part No.: 28-95-85, Sorin Group Deutschland Gmbh, Lindberghstrasse 25, Munchen, Germany, Sorin Group USA, Inc., 14401 W. 65th Way, Arvada, CO 80004. Used to perform, control and monitor extracorporeal blood circulation during an operation for up to six hours.

DateJan 26, 2011 (initiated Nov 12, 2010)
StatusTerminated
ReasonTouch screen may become unresponsive, inhibiting user input.
Root causeDevice Design
ActionAll affected US customers and distributors were notified by certified mail on Nov 12, 2010 via a Field Correction letter. They were told that they would be contacted by Sorin to arrange an appointment to inspect and replace affected touch screens on site. They were told that they could safely continue using their S5 systems in accordance with the Operator's Manual and previously supplied instructions until the inspection and necessary replacement had been completed. A customer response form was also included for completion and return to the firm. Customers can contact the firm if they have
Quantity2 unit
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060053STOCKERT S5 SYSTEM
Sorin Group Deutschland GmbH, Jun 2, 2006