Recall Z-1251-2011 ยท HAW Neurological Stereotaxic Instrument

Elekta, Inc.

Elekta Leksell Stereotactic System, used for localization (spatial reference) for cranial surgery using X-ray or CT and MRI Image data.

DateFeb 15, 2011 (initiated Oct 2, 2009)
StatusTerminated
ReasonNew protocols for MR sequences may result in higher RF energies deposited during MR scanning, generating heat in uninsulated fixation posts.
Root causeLabeling design
ActionElekta sent Important Notice Affected Product: Fixation Posts, A331, dated June 22, 2009. The Important Notice will be added to the Customer User manual upon receipt.
Quantity379 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K972324LEKSELL STEREOTACTIC SYSTEM
Elekta Instrument AB, Mar 2, 1998