Recall Z-1032-2011 ยท GEH Unit, Cryosurgical, Accessories

Galil Medical, Inc.

SeedNet MRI Cryosurgical System, model FPRCH2035, and the following system is not approved in the US: CryoHit MRI Cryoblation System, model FPRCH2048. The system is a software controlled cryoblation system designed to destroy targeted tissue using a minimally invasive percutaneous approach. Freezing and thawing of the tissue is achieved through the use of the Joule-Thompson effect, using a hig

DateFeb 3, 2011 (initiated Dec 20, 2010)
StatusTerminated
ReasonThis voluntary correction has been initiated due to the discovery of a damaged optocoupler on one of our Presice Systems during periodic preventive maintenance. The optocoupler sends a signal across the locking mechanism in the Presice and Seednet MRI Systems, enabling the software to recognize if the channel on the system manifold is locked or unlocked. If an optocoupler is damaged or malfunction
Root causeDevice Design
ActionGalil Medical sent an "Urgent Medical Device Correction" letter dated December, 30, 2010, to all affected consignees. The letter described the product, the problem involved, and the action to be taken by the consignee. Consignees were instructed to make sure the locking mechanism on the Precise and Seednet MRI Systems were in the "locked" position when a needle is inserted into the manifold. Gently pull on the black plug to ensure needle is secure in locking mechanism. Apply the enclosed alert sticker on the bottom of the distribution panel and the bottom of the system monitor of the Prec
Quantity22
DistributionWorldwide Distribution - USA including CA, CT, FL, GA, MA, MD, MN, MO, NC, NY, OH, PA, SC, TN, TX, VA, and WA and the countries of CANADA, DENMARK, FINLAND, FRANCE, GERMANY, HONG KONG, ISRAEL, ITALY, KOREA, NETHERLANDS, SPAIN, SWEDEN, TAIWAN, TURKEY, UNITED KINGDOM, CHINA, and JAPAN.CANADA, DENMARK,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021261SEEDNET SYSTEM AND SEEDNETGOLD SYSTEM, MODELS GP6T5, GP5T5, GP5T3, HP5T5 & HP5T3
Galil Medical , Ltd., Jul 3, 2002
K051052SEEDNET FAMILY (SEEDNET/SEEDGOLD SYSTEM, CRYOTHERA SYSTEM, CRYO-HIT SYSTEM)
Galil Medical , Ltd., May 6, 2005