Recall Z-1030-2011 ยท JAI Couch, Radiation Therapy, Powered

Elekta, Inc.

Precise Treatment Table The Precise Treatment Table is intended as a universal patient treatment support and positioning table for radiation therapy and simulation use. It is for use with existing Elekta Linear Accelerators and Simulators and adaptable to third party radiotherapy products.

DateFeb 3, 2011 (initiated Mar 10, 2008)
StatusTerminated
ReasonPossible error in the table longitudinal position.
Root causeOther
ActionElekta sent an updated IMPORTANT NOTICE letter dated December 24, 2008, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Elekta recommended that all users, and Eleckta field service engineers do regular QA checks to compare the movement of the treatment table in a given axis with the digital readout. Elekta Service representatives would attach the screws with Loctite and calibrate the axis on these sites. For questions regarding this recall call 770-670-2548.
Quantity597 units
DistributionNationwide Distribution and Puerto Rico
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K983678ELEKTA ONCOLOGY SYSTEMS PRECISE TREATMENT TABLE
Elekta Instruments, Inc., Jan 15, 1999