Recall Z-1548-2011 ยท MKJ Automated External Defibrillators (Non-Wearable)

Physio Control, Inc.

The LIFEPAK12 Defibrillator/Monitor. The LIFEPAK12 Defibrillator/Monitor can be used in AED mode, manual mode or advisory mode. In AED mode, the LIFEPAK 12 is intended for use by personnel who are authorized by a physician/medical director. In manual mode, the operator determines whether or not a defibrillation shock is needed by looking at the ECG rhythm display. Advisory mode can be useful durin

DateMar 4, 2011 (initiated Oct 1, 2010)
StatusTerminated
ReasonL1FEPAK 12 Operating Instructions that describe the process for disconnecting the device from AC (wall) power are not always being followed. When the instructions are not followed and the Power ON button is pushed within 2 seconds after disconnecting the device from AC power, the service indicator will illuminate and the device may fail to charge the defibrillator and/or initiate pacing.
Root causeDevice Design
ActionPhysio Control, Inc. began mailing the URGENT-MEDICAL DEVICE INFORMATION letter to their consignees on October 1, 2010. The letter identified the product, the problem, and the action to be taken by the consignee. The consignees were advised to follow the current LIFEPAK 12 Operating Instructions-Section 7.5-AC Power Adapter Operation and wait at least 2 seconds to push the Power ON button after disconnecting the device from AC power. If the instructions are not followed and the Power ON button is pushed within 2 seconds after disconnecting the device from AC power and the service indicato
Quantity39, 180 units (17168 units internationally and 22012 units in US)
DistributionWorldwide Distribution - USA and the countries of ARGENTINA, AUSTRALIA, BAHRAIN, BERMUDA, BOLIVIA, BRAZIL, BRUNEI, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, EGYPT, GERMANY, GREECE, HONG KONG, INDIA, ISRAEL, JAPAN, JORDAN, KENYA, KOREA, KUWAIT, LEBANON, MALAYSIA, MALTA, MEXICO, MOR
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010918LIFEPAK 12 DEFIBRILLATOR/MONITOR SYSTEM
Medtronic Physio-Control Corp., Aug 23, 2001