Recall Z-1459-2011 ยท MKT Hepatitis Viral B Dna Detection

Roche Molecular Systems, Inc.

COBAS AmpliPrep/COBAS TaqMan HBV v2.0 US-IVD; COBAS AmpliPrep/COBAS TaqMan HBV v2.0 CE-IVD

DateMar 1, 2011 (initiated Dec 28, 2010)
StatusTerminated
ReasonThere is an issue regarding liquid level detection of the CS1 Magnetic Glass Particles (MGP) Reagent Cassette of the COBAS AmpliPrep /COBAS TaqMan HBV v2.0 Test on the COBAS AmpliPrep instrument.
Root causeEmployee error
ActionRoche sent Urgent Medical Device Correction Letters to their consignees on December 28, 2010 via Fed Ex. The letters advise the customers to verify the correct configuration of the CSI reagent cassette, remind customers to follow the package insert instructions for including all required controls for each batch and advise that review of previously generated results may be warranted. A UMDC Faxback form was attached for the consignee to complete. For questions regarding this recall call 908-253-7569.
Quantity93 kits to US; 22,119 kits to foreign
DistributionWorldwide Distribution - USA including CA, CT, IN, MI, MN, NC, NE, NY, OR, TX, WA, and WI and the countries of Australia, Austria, Azerbaijan, Belgium, Brazil, Canada, China, Croatia, Czech Republic, Egypt, Finland, France, Germany, Ghana, Greece, Hong Kong, India, Israel, Italy, Japan, Kuwait, Luxe
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.