Recall Z-1019-2011 ยท LLZ System, Image Processing, Radiological
Philips Medical Systems North America Co. Phillips
PCR Eleva. The marketing brochure is labeled in parts: "Philips Medical Systems is part of Royal Philips Electronics... www.medical.philips.com... Philips Medical Systems Nederland B.V". Provides one or more capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images.
| Date | Feb 4, 2011 (initiated Dec 23, 2008) |
| Status | Terminated |
| Reason | There is a risk of selecting the wrong patient, when patient is selected in the patient list and the examination tab pressed with background query are running at the same time. |
| Root cause | Software design |
| Action | On 1/30/2009, Philips began sending out URGENT-Device Correction PCR Eleva notifications to their consignees. The consignees were informed of the hazards as when a patient is selected in patient list and examination tab pressed with background query running at the same time, there would be a risk that the wrong patient is selected. The firm issued the Field Corrective Order (FCO) 73200029 to their representative to correct the problem. The corrective action would be a software upgrade that resolve the issue. All consignees can contact Philips Call Center at 800-722-9377, #5, #2 and referen |
| Quantity | 686 units |
| Distribution | Nationwide Distribution -- USA, including states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WI, WV, and WY and to LANDSTUHL ARMY BASE, APO in United Arab Emirat |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K964124 | PHILIPS COMPUTED RADIOGRAPHY Philips Medical Systems, Inc., Dec 18, 1996 |