Recall Z-1651-2011 ยท OOI Real Time Nucleic Acid Amplification System
Roche Diagnostics Operations, Inc.
MagNA Pure LC 2.0 Instrument, catalog number 05197686001, Roche Diagnostics, Indianapolis, IN MagNA Pure LC 2.0 instrument is a robotic instrument for automated isolation of nucleic acids (DNA, total RNA , total viral nucleic acids) from different kinds of biological research sample material (whole blood, serum, blood cells, culture cells, tissue, bacteria, fungi) using the specially designed M
| Date | Mar 14, 2011 (initiated Oct 26, 2010) |
| Status | Terminated |
| Reason | If the bar code is printed from the Batch Results Screen, a wrong batch ID may be printed, while the correct batch ID is still assigned to the batch in the data repository of the MPLC 2.0 software. This may result in sample mis identification. If printed from the ordering screens, the bar code is printed with the correct batch ID. |
| Root cause | Software design |
| Action | The firm, Roche, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated November 04, 2010 to its customers. The letter described the product, problem and action to be taken. The customers were instructed to print batch ID barcode labels from the ordering or stage set up screens only; print Cooling Block barcode labels from the Post Elution Edit screen only; complete and return the attached UMDC Faxback Form for MagNA Pure LC 2.0 Instrument [UMDC 10-228] via fax to 1-877-551-7152 and file this correction notice for future reference. The firm plans to issue a software update in first quarte |
| Quantity | 119 devices |
| Distribution | Worldwide distribution: AL, AZ, CA, CO, DC, FL, GA, HI, IL, KY, LA, MA, MD, ME, MN, MO, MT, NC, ND NE, NH, NJ, NM, NY, OH, OK, OR, SC, SD, TN, TX, UT, VA, VT, WI and Puerto Rico; and country of Europe. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.