Recall Z-0931-2011 ยท FRN Pump, Infusion

Baxter Healthcare Corp.

Baxter Colleague Single Channel Volumetric Infusion Pump. Baxter Healthcare Corporation, Medication Delivery Division. Product Code: 2M8161

DateJan 21, 2011 (initiated Dec 1, 2010)
StatusTerminated
ReasonWhile the infusion pump was being serviced at the Baxter Service Center, it was inadvertently routed to the packaging area before the final inspection.
Root causeEmployee error
ActionBaxter telephoned the one customer on December 1, 2010, informing them that that an outgoing Quality Control Evaluation was not performed on their Colleague pump serial number 15030799CC during the last service. Since this was not conducted on this pump, the device may not function as designed and could lead to an interruption of therapy. Baxter requested that they locate the pump and remove it from use as soon as possible. Baxter will then send a shipping container for retrieval, reinspection and return of the device back to the customer at no cost. For questions regarding this recall call (8
Quantity1 pump
DistributionTexas
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063696COLLEAGUE VOLUMETRIC INFUSION PUMP MODEL# 2M8151,8153,8161,8163,9161,9163
Baxterhealthcare, Inc., Feb 27, 2007