Recall Z-1823-2011 ยท CIC Cresolphthalein Complexone, Calcium

Roche Diagnostics Operations, Inc.

Calcium Reagent, Calcium Test System for COBAS INTEGRA 400, 400 PLUS, 800, 800 CTS, cobas c 311, 501 and 502, Part number 20763128322, Roche Diagnostics Corporation, Indianapolis, IN. For the in vitro quantitative determination of calcium in serum, plasma and urine.

DateMar 25, 2011 (initiated Oct 11, 2010)
StatusTerminated
ReasonThe firm has recieved complaints of imprecision and erroneous patient results for calcium on the COBAS INTEGRA and COBAS c 501 system using reagent lots 62601901 and 63020401.
Root causeOther
ActionThe firm, Roche, sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated October 11, 2010 to all consignees/customers via UPS(signature required) . The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue use of the recalled product; discard from inventory according to their site's local regulations; complete and return the attached Faxback form via fax to 1-877-598-5705 even if they do not have the affected product, and file this letter for future reference. The firm added lot 63020401 to the recall in 3/2011. The firm is
Quantity13, 405 units
DistributionNationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K963292COBAS INTEGRA REAGENT CASSETTES & ANCILLARY REAGENTS
Roche Diagnostic Systems, Inc., Oct 31, 1996