Recall Z-1257-2011 ยท JAK System, X-Ray, Tomography, Computed

Toshiba American Medical Systems Inc

Toshiba TSX-301A AQUILION ONE (MPDCT0297EA) - Whole Body X-Ray CT Scanner

DateFeb 15, 2011 (initiated Sep 7, 2010)
StatusTerminated
ReasonIncorrect Dose Calculation. The software that performs the pre-scan dose calculation uses an incorrect parameter for an exam Plan using a tube voltage of 80kV or 100kV. This results in a pre-scan expected dose that is approximately 10% lower than what the displayed value should be when the scanner executes the exam Plan.
Root causeSoftware design
ActionToshiba America Medical Systems (TAMS) issued notification to consignees 9/7/2010, via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS as proof of notification.
Quantity44 units distributed between 09/2007 - 07/2010
DistributionNationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083282MODEL TSX-301B AQUILION CT SYSTEM
Toshiba America Medical Systems, In.C, Dec 12, 2008