ClearanceBrief
Recall Z-1248-2011 ยท EHD Unit, X-Ray, Extraoral With Timer

Gendex Dental Systems

Gendex Dental Systems, Gendex eXpert DC Intraoral X-Ray System; a dental x-ray system; Model 110-0205G1, catalog number EXPERTDC75NA

DateFeb 15, 2011 (initiated Dec 17, 2010)
StatusTerminated
ReasonThe intraoral x-ray systems were installed improperly with incorrect screws which allow the unit to pull away from the wall when the arm is fully extended. This condition may harm the patient.
Root causeComponent design/selection
ActionGendex sent a Urgent Medical Device Field Correction letter dated December 17, 2010 to the consignee, informing them that the x-ray units were installed with improper sized lag bolts and requesting that they contact Gendex Customer Service at 1-800-323-8029 to complete the field correction.
Quantity5 units
DistributionCalifornia
FDA recordOpen on fda.gov