ClearanceBrief
Recall Z-0784-2011 ยท JAK System, X-Ray, Tomography, Computed

GE Healthcare, LLC

HiSpeed FX/I (2201000-2 with Gantry mo. #2200997-2)

DateJan 4, 2011 (initiated Feb 1, 2010)
StatusTerminated
ReasonA defect with the hardware and software displaying the incorrect dose data.
Root causeRadiation Control for Health and Safety Act
ActionImportant Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to
Quantity102 total
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.