Recall Z-1590-2011 ยท GCJ Laparoscope, General & Plastic Surgery

Anchor Products Company, Inc.

Anchor Tissue Retrieval System; ETO sterilized, single use disposable device; port size 10 mm, 225 mL capacity; 3 units per box; Manufacturer: Anchor Products Co., 52 Official Rd, Addison, IL 60101 U.S.A.; Ref TRS100SB The device is a single use, sterile disposable pouch used with a dedicated introducer for the capture and removal of tissue or organ from the body cavity during laparoscopic surger

DateMar 7, 2011 (initiated Dec 9, 2010)
StatusTerminated
ReasonThe metal component within the unit can break and cause a malfunction. Use of the device may represent a potential health hazard if the metal component is not removed after breakage.
Root causeDevice Design
ActionThe firm, Anchor, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated December 8, 2010 to all customers. The letter described the product, problem and actions to be taken. The customers were instructed to examine their inventory and return any remaining units to their distributor for replacement; and complete and return the attached ANCHOR RECALL RETURN RESPONSE FORM via fax to (314) 961-4535 or mail to: Progressive Medical, Inc., Attn: Marketing Manager, 11085 Gravois Industrial Court, St. Louis, MO 63128, even if they have no inventory. A representative of Anchor Products then visited ea
Quantity88,785 units
DistributionWorldwide distribution: USA including state of: Missouri; and countries including: Australia, Belgium, Canada, France, Greece, Ireland, Japan, Spain and the United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061555ANCHOR LAPARSCOPIC TISSUE RETIEVAL SYSTEM
Anchor Products Co., Aug 31, 2006