ClearanceBrief
Recall Z-0875-2011 ยท DQX Wire, Guide, Catheter

Boston Scientific Corporation

Boston Scientific brand Synchro2 0.14" Guide Wire, 300 cm, Standard Pre-Shaped; Boston Scientific Corporation: Made in 2405 West Orton Circle, West Valley City, UT 84119 Product is distributed by Boston Scientific Corporation, Fremont, CA The Synchro guidewire series is intended for general intravascular use, including neurovascular and peripheral vasculatures. It can be used to selectively int

DateJan 7, 2011 (initiated Jun 7, 2010)
StatusTerminated
ReasonThe Guidewire has the potential to flake off polytetrafluroethylene (PTFE) particulate, which has the potential to embolize and result in infarction when introduced into a patient's circulatory system.
Root causeOther
ActionBoston Scientific Corporation sent an Urgent Medical Device Recall letter dated June 7, 2010, by Federal Express to all customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue use of and segregate the affected product. Complete and return the Reply Verification Tracking Form even if they did not have any affected product. Contact the Customer Service Call Center to obtain a Returned Goods Authorization (RGA) Number at (800) 811-3211, Option #1. E-mail or Fax the completed and signed Re
Quantity36, 049 total units, all sizes
DistributionWorldwide Distribution - USA including CA, CO, IN, MD, MI, OR, TX, and VA and the countries of Australia, Austria, Belgium, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Great Britian, Greece, Honduras, Hong Kong, India, Ireland, Isr
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K053268PVS 2000 SYNCHRO2 GUIDEWIRE
Boston Scientific - Precision Vascular, Mar 13, 2006