Recall Z-1807-2011 ยท KWA Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)
Depuy Orthopaedics, Inc.
ASR 300 Size 68, DEPUY ASR XL ACETABULAR CUP SYSTEM. The DePuy ASR XL Acetabular Cup components are part of a modular system designed to replace the natural articular surface of the hip joint in total hip replacement. The acetabular cup is designed as a cobalt-chrome-molybdenum (CoCrMo) alloy one piece cup with a porous coating and is available in outer diameter sizes 64mm through 70mm in two-mil
| Date | Mar 24, 2011 (initiated Aug 23, 2010) |
| Status | Terminated |
| Reason | As part of our ongoing post-market surveillance of all products, DePuy is continually evaluating data from a variety of sources including national joint replacement registries, published literature, company sponsored clinical trials, internal complaints data and unpublished clinical research reports. DePuy Orthopaedics issued a Field Safety Notice in March 2010 after receiving new data from the |
| Root cause | Device Design |
| Action | DePuy Orthopaedics issued a Field Safety Notice in March 2010 to all affected customers. The Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to share the Field Safety Notice with their organization and any organization where the ASR products may have been transferred. Customer were instructed not to implant the ASR devices. Patients who received the ASR System should be informed of this recall and instructed to return for a follow up visit. Patients with radiographic changes indicative of product failure should be addressed |
| Quantity | 46 |
| Distribution | Worldwide |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K080991 | DEPUY ASR XL MODULAR ACETABULAR CUP SYSTEM DePuy Orthopaedics, Inc., Jul 2, 2008 |