Recall Z-1971-2011 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

GE Healthcare, LLC

Precision 500D X-ray systems with under-table collimator part number 5234960. The Precision 500D is designed to perform radiographic and fluoroscopic x-ray examinations.

DateMay 6, 2011 (initiated Jun 10, 2010)
StatusTerminated
ReasonUse of high force when moving the image carriage/IDD into forward/park position can cause changes to the mechanical alignment of the collimator blades and allow the blades to open wider than specified, consequently allowing radiation outside the permitted field without warning to the user. This issue manifests with symptoms wherein the user cannot fully close down the lateral blades.
Root causeRadiation Control for Health and Safety Act
ActionGE Healthcare sent an Importan Electronic Product Radiation Warning letter to customers. The letter discussed the product defect description and related hazards, affected product details, instructions, and product correction. The firm will send a GE Field Service Representative to visit each affected customer site to inspect and correct all potentially affected units in the field following the instructions in field modification instructions No: 10846. If you have any questions please call 800-437-1171.
Quantity440 US
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.