ClearanceBrief
Recall Z-1571-2011 ยท JKA Tubes, Vials, Systems, Serum Separators, Blood Collection

Covidien LP

MONOJECT Blood Collection Tubes containing heparin 10x64 Sodium Heparin Item code 8881320256

DateMar 7, 2011 (initiated Nov 15, 2010)
StatusTerminated
ReasonThe heparin sodium USP used to manufacture blood collection tubes was produced with crude heparin containing Over-Sulfated Chondroitin Sulfate.
Root causeMaterial/Component Contamination
ActionCovidien sent an URGENT PRODUCT RECALL letter dated November 12, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately stop using the product, and return it to the firm. Distributors were instructed to notify their customers. All customers were required to respond to the notice regardless of their remaining inventory. Responses will be tracked for follow-up and effectiveness checks. For Customers Service questions regarding this recall call 1-800-962-9888 option 1.
Quantity126,000
DistributionWorldwide Distribution - USA including AL, AZ, CA, CO, DE, FL, GA, IA, ID, IL, KS, KY, MA, MI, MN, MO, MS, NC, ND, NE, NJ, NM, MV, MY, OH, OK, PA, SC, SD, TN, TX, WA, and WI and the country of Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K883320MONOJECT(R) TRACE ELEMENT BLOOD COLLECTION TUBE
Sherwood Medical Co., Sep 27, 1988