ClearanceBrief
Recall Z-0628-2011 ยท KOG

Advanced Sterilization Products

ASP Automatic Endoscope Reprocessor (AER), Product Code: 20300 & 20301 The AER is a machine designed to automatically wash and high-level disinfect flexible, submersible fiberoptic or video endoscopes.

DateDec 16, 2010 (initiated Feb 19, 2010)
StatusTerminated
ReasonThe recall was initiated because Advanced Sterilization Products (ASP) received three customer reports of elevated temperature of the high-level disinfectant solution used in the ASP AER as a result of a possible temperature monitoring system malfunction. This recall communication advised consignees that if high-level disinfectants are exposed to elevated temperatures, fumes may occur and resul
Root causePending
ActionAdvanced Sterilization Products sent an URGENT: PRODUCT CORRECTION letter dated February 19, 2010, to all customers. The letter identified the product, the problem, and the action to be taken by the customer. The recall communication was initiated with ASP forwarding a "Urgent: Product Correction" letter with attached customer fax-back response form to the affected customers. Customers were also instructed to complete and sign the enclosed customer fax-back response form indicating that they have disconnected the heater on the ASP AER according to the instructions provided and fax the fo
Quantity1,881 units (1514 in US)
DistributionWorldwide Distribution -- United States, Argentina Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, Hungary, Mexico, Panama, Guatemala, Honduras, Poland, Puerto Rico, Virgin Islands, Singapore, Taiwan, Venezuela, Croatia, and Bulgaria.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K832458AUTOMATIC ENDOSCOPE
American Endoscopy, Inc., Oct 20, 1983