Recall Z-1011-2011 ยท CEM Electrode, Ion Specific, Potassium

Radiometer America Inc

ABL90 FLEX Blood Analyzer Part Number: 393-090. Intended for use by trained technologists, nurses, physicians and therapists and for use in a laboratory environment, near patient or point-of-care setting. Also, Intended for in vitro testing of samples of heparinized whole blood.

DateFeb 2, 2011 (initiated Nov 1, 2010)
StatusTerminated
ReasonDue to a software error, a patient mix up can occur on the ABL90. If a sample is pre-registered and the sample "fails" the next result will inherit the patient data from the previous result.
Root causeSoftware design
ActionOn 11/16/10 the firm sent Medical Device Field Correction letters to their customers. The letter states that all systems will be upgraded and until each system is upgraded, they instructed their customer to "always query for patient ID prior to introducing the sample." They also instruct each customer to complete the "Fax Return Form" acknowledging receipt of the letter. If customers have questions regarding the letter or the upgrade, they should contact Technical Support at 1-800-736-0600 Opt 4.
Quantity432 units
DistributionWorldwide Distribution -- USA, including states of MI, NJ, & PA and countries Australia, Austria, Belgium, Denmark, Finland, France, Germany, Indonesia, Italy, Japan, Kuwait, Malaysia, Netherlands, Norway, Poland, Russia, Spain, Sweden, Switzerland, United Arab Emirates & United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092686ABL90 FLEX SERIES
Radiometer Medical Aps, Aug 6, 2010