ClearanceBrief
Recall Z-1375-2011 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Healthcare, LLC

GE Healthcare Discovery MR750 3.0T The GE Signa MR750 System is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The GE Signa MR750 System is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of t

DateFeb 18, 2011 (initiated Nov 18, 2010)
StatusTerminated
ReasonGE Healthcare has become aware that during certain service procedures, there is a potential for electric shock associated with the gradient cooling manifold hoses of your MR750, MR450, and MR450w MR systems that may impact service or maintenance personnels safety. The gradient cooling manifolds used on the affected systems may contain electrically conductive hoses, which become live when the g
Root causeNonconforming Material/Component
ActionGE Healthcare sent an "Urgent Medical Device Correction" letter dated November 16, 2011, to all affected customers. The letter was addressed to Hospital Administrator / Risk Manger, Radiology Department Manager and Maintenance / Biomedical Engineering Department. The letter described the Safety Issue, Affected Product Details, Safety Instructions, Product correction and Contact Information. Customers were instructed to follow appropriate lockout/tagout (LOTO) procedures. Customers were instructed to contact GE Healthcare engineering at 843-687-7213 or 843-697-3009 for instructions on how to p
Quantity25
DistributionWorldwide Distribution - USA including AL, AZ, CA, CO, CT, DE, DC, FL, GA, IL, IN, IA, LA, MD, MA, MI, MO, NV, NJ, NY, NC, OR, PA, SC, TN, TX, WA, WI. and the countries of AUSTRALIA, BRAZIL, CANADA, EHYPT, FINLAND, FRANCE, GERMANY, GREECE, INDIA, ITALY, JAPAN, KUWAIT, QATAR, REPUBLIC OF KOREA, SWIT
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081028SIGNA 3.0T MR750 SYSTEM
Ge Medical Systems, LLC, Apr 25, 2008