ClearanceBrief
Recall Z-1389-2011 ยท LXH Orthopedic Manual Surgical Instrument

Depuy Orthopaedics, Inc.

S-ROM Distal Femoral Cutting block Guide SZ MED, REF 216311004, non-sterile, Depuy International Leeds, England. The product is used to make cuts in the femur to prepare the femur for total knee replacement surgery.

DateFeb 22, 2011 (initiated Oct 19, 2010)
StatusTerminated
ReasonThe firm was notified that the Hinge Block Femoral Box Guide cut being off alignment when using the S-ROM Distal Femoral Cutting Block Guide SZ SML/X-SML pin holes to locate. The firm then found that the same issue was found int the MED block size.
Root causeDevice Design
ActionThe firm, DePuy, sent an "URGENT INFORMATION - DEVICE CORRECTION" email and letter dated October 19, 2010 to all customers. The notices described the product, problem and actions to be taken by the customers. The notices stated that: Depuy is issuing a device correction; however, the product may still be used; the devices do not need to be returned; the distributors were to provide a letter to clinicians that were using the device; the surgical technique will be revised in October 2010; the product will undergo a design change, and the recalled product will be traded out in December 2010 a
Quantity219
DistributionWorldwide distribution: USA including states of: AZ, AR, CA, CO, FL, GA, IA, ID,IL, IN, KY, KS, LA, MA, MD, MI, MN, NC, NV, PA, TN, TX, VA, WA, and WI; and countries including: Canada and Ireland. Australia, Austria, Canada, Denmark, Ireland, Israel Italy, New Zealand, Russia, Sweden, Switzerland, S
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.