ClearanceBrief
Recall Z-0562-2011 ยท JOR Generator, Pulsatile Flow, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corporation

Sarns 8000 Roller pump for Sarns System 8000 Heart Lung Machine console, 115V, Catalog No 16402, Terumo Cardiovascular System Corp Ann Arbor, MI. The system 8000 is indicated for use in extracorporeal circulation of blood for arterial perfusion, reginal perfusion and cardiopulmonary bypass procedures.

DateDec 8, 2010 (initiated Nov 2, 2010)
StatusTerminated
ReasonThe system 8000 roller pumps may fail resulting in a persistent pump stop.
Root causeOther
ActionThe firm sent "URGENT MEDICAL DEVICE CORRECTION" dated 11/02/2010 to customers. The letter describes the product, problem and corrective actions taken and/or needed to be taken by the customers. The letter states the system responds to a failed inductor by displaying an overspeed message and stopping the roller pump without an audible alert. The pump cannot be restarted. The customers/users are instructed to stop the pump, replace the pump with a backup or if necessary use the handcrank. The letter states that Terumo will replace the roller pump at the next preventative maintenance visit
Quantity95 roller pumps.
DistributionWorldwide distribution: USA including states of CA, FL, NC and NY and countries of United Arab Emirates, China, Ecuador, and Panama.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K915183SARNS 8000 MODULAR PERFUSION SYSTEM
3M Health Care, Sarns, Jul 8, 1992