Recall Z-0915-2011 ยท GXD Generator, Lesion, Radiofrequency

Stryker Instruments Div. of Stryker Corporation

Stryker Instruments, Interventional Spine RF MultiGen REF 0406-900-000, Rx only,, Stryker Instruments, Kalamazoo, MI (USA), 49001, (269) 323-7700, (800) 253-3210. The product is intended for coagulation of soft tissues in orthopedic, spinal and neuorsurgical applications.

DateJan 18, 2011 (initiated Apr 13, 2010)
StatusTerminated
ReasonThis recall is related to the Electri-cord recall. There is a risk when excessive force is applied to the plugs, a fracture can develop on the prongs inside the molded section of the plug. Over time the fracture may lead to arcing inside of the molded section of the plug and a build up of heat which can cause charring, electric shock or fire. To this date there have been zero complaint for this
Root causeComponent design/selection
ActionStryker Instruments sent an Urgent - Medical Device Recall Notification letter dated April 13, 2010, to all affected customers. The letter identified the product, the problem,and the action to be taken by the customer. Stryker notified customers again on May 24, 2010, and again on September 14, 2010, via email, fax, or telephone. Customers were instructed to locate the units and visually inspect each affected cord. Remove the affected cord and dispose of it using their healthcare facilities disposal procedure. If the cord did not show signs of damage, the unit could remain in service until
Quantity3364 all products
DistributionWorldwide Distribution - USA (nationwide) and the countries of Portugal, Canada, China, Taiwan, France, Germany, Greece, Spain, India, Italy, Korea, Switzerland, Singapore, and England.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032601STRYKER INTERVENTIONAL PAIN RF GENERATOR
Valley Forge Scientific Corp., Apr 1, 2004