Recall Z-0797-2011 ยท LLZ System, Image Processing, Radiological
Vital Images, Inc.
Vital, Vitrea fX and Vitrea Enterprise Suite fX Model number: Vitrea fX 2.1, fX 3.0, and fX 3.1 Vitrea Enterprise Suite (VES) fX 1.2 and VES fX 1.3 The Vitrea 4D CT Brain Perfusion option is intended for post processing based on dynamic CT images continuously acquired during the infection of contrast, for the visualization of apparent blood flow in brain tissue and pictorial illustration of perf
| Date | Dec 21, 2010 (initiated Oct 14, 2010) |
| Status | Terminated |
| Reason | Vital Images Inc., reported that a software defect exists with Vitrea fX 2.1, 3.0 and 3.1 CT Brain Analysis whereby the annotated numeric ROI values on a batched perfusion map are incorrect when created using the auto batching function (from the right click menu) with the auto ROI templates enabled. |
| Root cause | Software design |
| Action | Vital Images, Inc sent a Vital "Urgent Software Update Alert" letter dated 14 October 2010 to all Consignees/Customers. The letter described the product and problem. They advised, until new software is installed, that the customers can avoid obtaining incorrect values by tuning off ROI templates prior to selecting the batch option from the right click menu in the CT Brain Analysis application; and recommended that the customers re-examine any ROI perfusion values that were previously generated using automatic collage batching with ROI templates. Please direct any questions or comments to Cu |
| Quantity | 182 |
| Distribution | Worldwide distribution: USA including states of: AL, AZ, AR, CA, CO, FL, GA, IA, KY, LA, MD, MA, MI, MN, MO, MT, NY, NJ, ND, OH, PA, PR, TN, TX, WA, WI, and WY, and countries of: CANADA, NETHERLANDS, TAIWAN, CHINA, SINGAPORE, UNITED KINGDOM, BRAZIL, GERMANY, RUSSIA, THAILAND, FRANCE, REPUBLIC OF KOR |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K072821 | VITREA 4DCT Vital Images, Inc., Feb 20, 2008 |