Recall Z-0615-2011 ยท GEH Unit, Cryosurgical, Accessories

Endocare, Division Of Healthtronics

Incorrect Label: Endocare 2.4mm VARIABLE ATTACHED CRYOPROBE, REF: CVA 2400, LOT 2751. Correct Label: Endocare 2.4mm CRYOPROBE, REF: CRYO-44F (Sharp Tip).

DateDec 13, 2010 (initiated Nov 4, 2010)
StatusTerminated
ReasonMislabeled : The label on the outer box and inner pouch of the EndoCare 2.4mm CryoProbe (CRYO-44F) inaccurately reference the CVA2400 (2.4mm Variable Probe aka: V-Probe).
Root causeOther
ActionHealthronics issued an Urgent Device Recall letter dated November 2, 2010 to consignees. The letter described the product mislabeling, and requested customers to examine their inventory, quarantine all affected product, and complete and return the verfication section of the letter to the firm. A return authorization number and prepaid return label were provide to facilitate return and replacement. Healthronics can be contacted at 877 367-4873 concerning this recall.
Quantity33 Units
DistributionNationwide Distribution: to the states of FL, MI, PA, TN, and WA.