Recall Z-0802-2011 ยท FQO Table, Operating-Room, Ac-Powered

Kreuzer Gmbh & Co. KG

klinoPORT, AC powered Operating Table

DateDec 22, 2010 (initiated Oct 26, 2010)
StatusTerminated
ReasonThe motorized adjustable pendant arm may suddenly descend to its lowest position.
Root causeOther
ActionConsignees were notified by a Warning Notice on/about 10/26/2010. Medical facilities were instructed to adjust the pendant arm to the lowest height position and disable the motor and to place the accompanying warning label on the pendant. They will, therefore, not be able to adjust the height of the pendant until corrective action is developed and completed. TRUMPF Medical Systems, Inc. will receive corrective action repair kits by 11/12/2010.
Quantity15 units
DistributionNationwide Distribution -- CT, MA, and MO.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.