ClearanceBrief
Recall Z-0944-2011 ยท JAK System, X-Ray, Tomography, Computed

GE Healthcare, LLC

GE Healthcare STe (16 slice) System; Discovery STe (16 Slice) Systems with technical Reference Manual numbers 2351785-xxx and 2390935-xxx distributed from November 2001 to August 2010. Head and whole body X-ray computed tomography applications.

DateFeb 7, 2011 (initiated Oct 27, 2010)
StatusTerminated
ReasonAlthough the system x-ray output is accurate for the technique factors given, and the dose display is also accurate, certain CTDI adjustment factors used for the manual calculation from the Technical Reference Manual (TRM) are incorrect. CTDI100 Aperture Adjustment Factors for 4x0.625 mode contain incorrect values. This issue may lead to inaccurate CTDI manual calculation. The results of the diffe
Root causeRadiation Control for Health and Safety Act
ActionAn Important Electronic Product Radiation Warning letter will be provided together with the TRM addendum informing the customers of the issue and related hazards. The letter also identified the affected product and gave instructions, which include referring to the TRM addendum, along with the product correction. This correction will be executed via a Field Modification Instruction IFMI 254281 released in November, 2010 with an estimated completion date of May, 2011. Questions or concerns should be directed to the call numbers listed or a local GE Healthcare Service Representative for countr
Quantity1450 units total in US
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.