Recall Z-1943-2011 ยท DTY Sensor, Blood-Gas, In-Line, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corporation

Terumo Cardiovascular Systems Corporations, CDI 101 Hematocrit/Oxygen Saturation Monitoring System. In-line blood-gas sensor used in cardiopulmonary bypass.

DateApr 11, 2011 (initiated Nov 1, 2010)
StatusTerminated
ReasonInaccurate result. Possibility for displayed values for potassium, CO2, hematocrit, O2 saturation, pH, hemoglobin and other measured blood parameters to be inaccurate.
Root causeDevice Design
ActionOn 11/1/2010, Terumo sent "URGENT MEDICAL DEVICE CORRECTION" notices to Chief Perfusionist, Department of Cardiovascular Surgery or Director of Operating Room Services. The letter informed the consignees about the incidents of the CDI 100 and 101 displaying inaccurate values. The letter states that when a displayed value is significantly different from expectations based on the clinical situation, the user should verify the accuracy by independent means before treating. The letter included "Caution" stickers that were to be applied to the CDI system operators manual.
Quantity468 monitors
DistributionUS, India, Colombia, Taiwan, Venezuela, Vietnam, Puerto Rico, Malaysia, Sri Lanka, Brazil, Indonesia, Pakistan, Chile, Germany, Hong Kong, Japan, AE, Singapore, Austria, Belgium and Thailand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K902654CDI (TM) SYSTEM 100 MONITORING SYSTEM
3M Health Care, Ltd., Nov 5, 1990