ClearanceBrief
Recall Z-0557-2011 ยท DQO Catheter, Intravascular, Diagnostic

Merit Medical Systems, Inc.

Custom Kit, Sterile EO, Rx Only, Catalog No. K09-09815, Manufacturer: Merit Medical Systems, Inc. South Jordan, Utah 84095, USA,

DateDec 8, 2010 (initiated Oct 25, 2010)
StatusTerminated
ReasonDefective custom kits may draw in air during use which may compromise patient safety and affect treatment. The valve's fixed luer may be over- or under-tightened, causing leaking at the bond joint.
Root causeProcess control
ActionMerit issued an Urgent Product Recall Notice letter dated October 25, 2010 to consignees. Copnsignees were instructed to quarantine the product immediately for return and to notify customers of recall activity and instruct them to return product.. All distributors and customers are to complete and return a Recall Notification Form. For further information about this recall, contact Merit at 800 356-3748.
Quantity1033 units
DistributionNationwide Distribution in the states of CA, IL, IN, and UT.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K913682MERIT CUSTOM KIT
Merit Medical Systems, Inc., Nov 15, 1991