Recall Z-0428-2011 ยท NRY Catheter, Thrombus Retriever

Penumbra Inc.

Penumbra System Reperfusion Catheter 032, catalog number PSC032, lot number lot F15020, manufactured by Penumbra, Alameda, CA.

DateNov 24, 2010 (initiated Oct 25, 2010)
StatusTerminated
ReasonReperfusion catheter may fail at the mid-shaft joint and may break off. If the segment cannot be removed, it may result in blocked blood flow, requiring emergency medical treatment.
Root causeProcess control
ActionPenumbra issued an Urgent Voluntary Field Removal letter dated October 28, 2010 to consignees, identifying the affected device and actions to be taken by consignees. Customers were instructed to inspect their inventory for the affected product and quarantine it, pending return to Penumbra. Penumbra employees will contact customers to arrange return and replacement of affected units. Customers were also requested to complete and return a product identification form and return form to Penumbra within 5 business days. Customers can contact Penumbra at 1 510 748-3223 concerning this action.
Quantity61
DistributionWorldwide Distribution: 17 consignees in the US, including in the states of CA, FL, HI, ID, IL, KY, MA, MS, NJ, NY, PA, and TX, 1 in Canada and 21 in the EU countries of France, Germany, Hungary, Italy, Norway, Slovenia, Spain, and the United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072718PENUMBRA SYSTEM
Penumbra, Inc., Dec 28, 2007