Recall Z-1250-2011 ยท LLZ System, Image Processing, Radiological

GE Healthcare, LLC

Xeleris Processing and Review Workstation 1.0, 1.1, 2.0, 3.0 The display, processing, archiving, and communication of data acquired by Emission Tomography cameras used in diagnostic radiology, including procedures for planar imaging, whole body imaging, tomographic (SPECT) imaging, positron imaging by coincidence, attenuation correction, and anatomical image registration. 3X The system is intende

DateFeb 15, 2011 (initiated Jan 12, 2011)
StatusTerminated
ReasonGE Healthcare has become aware of safety issues relevant for the Xeleris Workstation 1.0, 1.1, 2.0, 2.1, 3.0: 1. Measurements taken on the CT Images on the CT system are different from the Xeleris Workstation 2. Inaccurate PET SUV values may be calculated when data is retrieved from PACS after being processed in the Xeleris. 3. Wrong marking of Early and Late result series on display in Bra
Root causeSoftware design
ActionThe firm, GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated January 12, 2011 to all consignees/customers. The letter described the product, problem and actions to be taken by the customers. The customers were instructed to please ensure that all potential users in their facility are made aware of this safety notification and the recommended actions. A GE Healthcare Service representative will perform the required software update on each affected system. If you have any questions or concerns regarding this notification, contact your local GE Healthcare Service Representa
Quantity6046
DistributionWorldwide distribution: USA including DC, PR, and Guam; and countries including: AUSTRIA, AUSTRALIA, ARGENTINA, ALGERIA, AZERBAIJAN, BELGIUM, BOSNIA & HERZEGOVIA, BULGARIA, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, FINLAND, FRANCE, GERMAN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K024137JUPITER PROCESSING AND REVIEW WORKSTATION
Ge Medical Systems, Inc., Jan 15, 2003
K051673XELERIS 2 PROCESSING AND REVIEW WORKSTATION
GE Medical Systems, Jul 18, 2005
K093982XELERIS 3 PROCESSING AND REVIEW WORKSTATION
Ge Medical Systems F.I. Haifa, Jan 8, 2010